3.2 Installation Space
4. Check against Packing List
For composing a liquid chromatograph system with the
Chromaster series module units, a space of 540 mm wide
(980 mm for stacking the units in two blocks) and 550 mm deep
is required at least. Behind each unit, a space of 150 mm or
more must be secured. At the left and right sides of the
respective units, a space of minimum 100 mm each is necessary.
For installation, prepare a level and sturdy bench which
withstands a load of 150 kg or more. The system conceived
here is an example in which Chromaster organizer, Chromaster
5110/5160 pump, Chromaster 5210/5260 autosampler,
Chromaster 5310 column oven and Chromaster
5410/5420/5430/5450 detector are stacked. Confirm the
necessary space taking the actual system configuration and OA
equipment in the installation room into account and secure an
adequate space at the time of installation.
Check the delivered items against the packing list.
APPENDIX 1 - 7