(4)
Operation Check
Plug the power cord into the instrument and turn on the
power switch. Make sure the lamp comes on.
NOTE: The D
lamp does not come on if it has been
2
mounted improperly.
(5)
On the monitor screen 2, check the wavelength (WL) and
reference energy (REF), and write down their values as a
record of the D
2
as a reference for the next D
historical record of D
Monitor Screen 2
WL
SAMP
250
191000
(6)
Enter the lamp replacement date as follows.
GLP
(a) Press
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
LAMP SELECT (D2=1, W=2, Hg=3)
1
(c) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
LOGBOOK (REPORT=1, RESET=2)
1
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
D2 LAMP CHANGE (MM DD YYYY)
<1–12>
5 - 7
lamp after use. This record can be used
lamp replacement and as a
2
lamp replacement.
2
REF
CELL
202000
ON
key, and the following screen will appear.
_9
15
2010
5.2.1
[I]