Hitachi CHROMASTER 5160 Instruction Manual page 601

Table of Contents

Advertisement

• Assembly of Constant Temperature Flow Cell (Option)
Chromaster 5440 fluorescence detector can incorporate the
constant as an option.
The temperature switching equipment installed in the constant
temperature flow cell keeps constant the temperature of the
mobile phase and the sample which is poured into the flow cell.
Therefore, change in sensitivity caused by environment
temperature change can be reduced.
NOTE: It is not permitted for the user to carry out
installation at the time of delivery. For ensuring
safety and high accuracy in operation of the
instrument, installation of the instrument shall be
carried out by qualified service personnel trained
by the manufacturer or its authorized
maintenance service agent.
(1)
Turn off the power to the instrument.
(2)
Remove the inlet tube connected to the bulkhead union of
the standard flow cell.
(3)
Loosen the retention screws to remove the standard flow
cell.
(4)
Connect the constant temperature flow cell connector.
(5)
Install the constant temperature flow cell, and fasten it by
tightening the retaining screws.
(6)
Set up the constant temperature flow cell setting as use
(Refer to section 3.5.2).
APPENDIX 1 - 16