No.
Check Item
6
Wash
mechanism
Procedure
Check the amount and generated air
bubbles of wash fluid in the wash
mechanism every day.
Depends on the state of the wash
fluid, it causes deterioration of the
pump seal and the plunger.
Leaks lead to poor baseline stability
and data reproducibility. The net
result of leaks is degradation of
reliability in measurement.
4 - 10
Remarks