3.2.5 Pressure Calibration Setting

Set up a calibration coefficient for the pressure.
Monitor screen
SET UP
UTLTY
UTLITY
(CONDITION=1, UNIT=2,
1
EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
Select CONDITION
1
ENT
CONDITION SET
4
PURGE=3,
Select CALIBRATION
4
ENT
CALIBRATION
(FLOW=1,
2
Select PRESS
2
ENT
PRESSURE CALIB
1.000
Enter calibration coefficient.
ENT
Monitor screen
3 - 22
FLUSH=3
(Z-ADJ=1,
MOTOR-CTRL=2
CALIBRATION=4,
PRESS=2)
(0.900 – 1.100)
LPGR=5)