3
Connection and Assembly
Additional DC mode connection (optional)
Connecting load lines
Connecting sense lines
Using an emulator adapter with a Scienlab Charging Discovery System
32/51
The function in DC mode can be added in addition to the AC operating mode.
The cables to the load connections and to the voltage measurement of the
DUT are routed via a replaceable cable gland plate underneath the output
box of the SL1200A, and connected via the corresponding terminal strips.
The cables must be connected to the DUT.
Refer to the enclosed circuit diagram for the detailed terminal assignment.
If the SL1200A Scienlab Regenerative AC Emulator is used together with a
Scienlab Charging Discovery System (CDS: SL1040A-ST2 or SL1047A) for
charging technology application, instead of load and sense lines, an emulator
adapter is connected to the Scienlab Regenerative AC Emulator, which is
permanently installed with the output box and no connection inside the
outbox is needed.
Scienlab Regenerative AC Emulator – SL1200A Series – Operating Instructions
Connection of DC LV load lines (SL1201 – SL1203)
• Connect to the DC+/DC- terminal strip according to circuit
diagram of the SL1200A
• From left to right: DC+/DC-
• Separate bolt for PE
• Ring cable lug connections
• Tightening torque: see Installation Guide
Connection of DC HV load lines (SL1212, SL1213)
• Connection according to circuit diagram of the SL1200A
• From left to right: DC+/DC-
• Separate bolt for PE
• Ring cable lug connections
• Tightening torque: see Installation Guide
Connection of DC sense lines, HV + LV
• Connect to the fuses according to circuit diagram
• Measure DUT voltage Usense
• From left to right: DC+/DC-
• Wire end ferrules connections