4
Commissioning
DANGER
CAUTION
NOTICE
NOTICE
Scienlab Regenerative AC Emulator – SL1200A Series – Operating Instructions
Danger to life in case of misuse at the outputs.
The Scienlab Regenerative AC Emulator does not have its own insulation guard.
If the system emulates an electrical isolation against PE at the output, the local
installation regulations and guidelines must be observed.
This would be in the case of the emulation of an IT network in AC mode, or
generally in DC mode.
Electrical malfunction through moisture condensation!
In case of high fluctuations of temperature, condensate water can be formed. Thus,
possible insulation sections cannot be maintained, and the electronics are damaged.
Before switching on the system, the Scienlab Regenerative AC Emulator
acclimatize to the ambient temperature in the operating area (generally between
5 °C and 40 °C) for at least 24 hours.
Hearing protection due to noise exposure in the operating area!
Noise level <85 dB(A) at a distance of 1 m from the front
It is recommended to wear hearing protection during operation.
The system may only be switched on and put into operation by specialist staff and
specially trained personnel. (see
The following cabling must be completed before commissioning:
Electrical installations are complete (see
Load and measuring lines are connected with the DUT (see
DUT),
Signal and control lines are connected (see
Control
Lines),
All connections and cabling between SL1200A, control computer /
Hardware-in-the-Loop (HiL) platform and higher-level safety technology, if
available, have been checked and are according to the enclosed circuit
diagram,
the device under test is in a protected environment, in accordance to the
safety guidelines, especially regarding protection earthing;
all necessary enable signals are provided,
PC software and firmware, available via Keysight Homepage
(Software/Firmware), are installed.
Qualification of the Operating
Electrical
Connecting Signal Lines and
Commissioning
Personnel)
Installation),
Connection of
39/51