Troubleshooting And Fault Elimination; Fault Message Via The Pc Software; System Cannot Be Switched On - Keysight SL1200A Series Operating Instructions Manual

Scienlab regenerative ac emulator
Hide thumbs Also See for SL1200A Series:
Table of Contents

Advertisement

6

Troubleshooting and Fault Elimination

NOTICE

Fault message via the PC software

System cannot be switched on

Scienlab Regenerative AC Emulator – SL1200A Series – Operating Instructions
Troubleshooting and fault elimination may only be carried out by specially trained
personnel. (see
Qualification of the Operating
The following overview provides information about faults, their causes and
remedies.
If an overload or a fault current interruption occurs or if a fault condition
exists:
Notify trained service personnel.
If necessary, notify KeysightCare
Contact KeysightCare
Keysight Technologies Technical Support
If the signal lamp or PC software
SL1200A Regenerative AC Emulator Soft Front Panel (SFP)
report an error, the error description can be read out via the PC software. The
detailed user manual is included in the scope of delivery.
The system does not switch on:
Possible causes
No supply voltage present
Emergency stop push button pressed
Signal via higher-level emergency stop/release
Circuit breaker
Fuse has tripped
Not properly wired external emergency/release
in/out
Connectivity problem
When option Increased Safety is available:
connectors defect
Table 5: Possible cause and remedy
Troubleshooting and Fault Elimination
Personnel)
www.keysight.com/find/keysightcare-contact
Remedy
Switch on the building supply voltage
Release emergency stop push button
Check laboratory safety technology
Remove front panel, switch to ON
Switch fuse on after opening the panel; reattach
panel afterwards
Check wiring inside the SL1200A
Check cabling
Contact KeysightCare