Data Record Format Examples - Hach Met One 6000 Series User Manual

Table of Contents

Advertisement

FXB communication
Information
Flow rate
Calibration value
Air temperature
Relative humidity
B.3.1 Data record format examples
2-channel with flow measurement example:
032408 080715 0100 0.5 000278 5.0 000013 CAL 000100 FLO 000100 LOC 000001 C/S 001512
2-channel with no flow measurement example:
032408 080717 0100 0.5 000278 5.0 000013 CAL 000100 LOC 000003 C/S 00155A
2-channel enviro probe and flow measurement example:
032408 080712 0100 0.5 000278 5.0 000013 R/H 000561 TMP 000231 CAL 000100 FLO 000100 LOC 000002 C/S 0016B1
68
Table 34 Data record element descriptions (continued)
Flow rate value shown in CFM. A value of 000100 equals .100 CFM. A value of 001000
equals 1.000 CFM.
Calibration value of the sensor. A value of 000100 equals a calibration voltage of 1.00 VDC.
Valid ranges are 0.80 to 1.20 VDC. A reported value outside this range will cause a
calibration alarm to be reported in the status byte.
Air temperature reading from the externally attached environmental probe. A value of
000231 equals 23.1°C.
Relative humidity reading from the externally attached environmental probe. A value of
000561 equals 56.1% RH.
Refer to
Figure 29 on page 66
Description
and
Table 34 on page 67
for element descriptions.