Utility Program Operation - Hach Met One 6000 Series User Manual

Table of Contents

Advertisement

Operation

4.1.2 Utility program operation

30
Complete the following steps to configure the particle counter.
1. Open the SetupUtility.exe file. The utility program will open (refer to
2. Find the Comport field. If necessary, change the COM port to match the port on the
PC that the particle counter is connected to. Click
read the data that is stored in the instrument.
3. Make sure that the data in the Instrument Information section is accurate. This
section shows the instrument model number, communication option, firmware version
and communication address (if applicable).
4. Change the parameters in the General section as is necessary. Refer to the
parameter descriptions that follow:
Count Mode (for pulse, Modbus or FXB only)—set to differential or cumulative
(refer to
section 4.2.3 on page
System Date/Time—enter the current date (YYYY/MM/DD) and time
(HH:MM:SS, 24-hour format).
Sample Timing: Sample—the length of time that data is collected for each
sample. The default sample time is one minute (00:01:00).
Sample Timing: Hold—the length of time that data collection is stopped between
samples. Use the Hold time to stop data collection during maintenance
procedures. The default Hold time is 0 (00:00:00).
Count Alarms—set the number of counts for each size channel that will trigger a
count alarm.
Note: The pulse unit uses only 2 channels for particle counts.
Sample Mode—set to Auto.
Flow Units—set to CFM (cubic feet per minute) or LPM (liters per minute).
5. The Diagnostics section can be used to make sure that the wiring to an external
light stack is correct. Set the Indicator LED to flash or not flash for one of the colors.
Look for the LED on the light stack to illuminate or flash to make sure that the wiring
is correct.
Note: It is not possible to save the diagnostic settings and they have no effect on the instrument
operation. For a description of the LED indicators, refer to
6. Change the settings for the communication protocol that is used:
Serial—select the RS485 serial communication protocol (FXB or Modbus). If
Modbus is selected, enter the slave address. When the address is 31 or less,
use the dip switches on the bottom of the instrument to set the address (refer to
Table 8 on page
Note: If an address of 32 or higher is entered, the dip switch setting will be ignored and the
entered value will be used.
Analog—set the count value for each channel that corresponds to the 20 mA
output signal (default = 1000).
Pulse—channel 1 is always set for the smallest particle size. Select a channel to
correspond to the largest particle size.
Ethernet—refer to
Wireless—refer to
7. Click
SAVE SETTINGS
36). The default count mode is set as cumulative.
33).
section 4.2.2 on page
section 4.3 on page
37.
to save the settings in the instrument.
Figure
. The utility will
READ INSTRUMENT
section 2.4 on page
34.
21).
7.

Advertisement

Table of Contents
loading

Table of Contents