A.3 Data label
Address
Register description
200
Size 1 label
202
Size 2 label
204
Size 3 label
206
Size 4 label
208–231
Reserved
232
Analog input 1 label
234
Analog input 2 label
236
Analog input 3 label
238
Analog input 4 label
240–251
Reserved
252–299
Expansion
A.4 Sample data
Address
Register description
300
Sample UTC timestamp—year
301
Sample UTC timestamp—month/day
302
Sample UTC timestamp—hour
303
Sample UTC timestamp—minute/second
304
Sample period—hours
305
Sample period—minutes/seconds
306
Reserved
307-308
Sample volume
309
Sample status. Refer to
310
Reserved
311
Size 1 counts
313
Size 2 counts
315
Size 3 counts
317
Size 4 counts
319–342
Reserved
343
Analog channel 1 (flow)
344
Analog channel 2 (temperature)
345
Analog channel 3 (relative humidity)
346
Analog channel 4
347
Analog channel 5 (CAL)
348–352
Reserved
353
Location name
Table 18
provides a register for sample and analog data labels.
Table 18 Count bin data labels
Sample data records
(Table
sample and hold times in the configuration registers. If real-time data is not required, use
the buffered records (address 500+).
Table 19 Sample data
Table
20.
Access
Size (bytes) Data format
R/P
4
R/P
4
R/P
4
R/P
4
R
4
R
4
R
4
R
4
19) are updated at each polled interval regardless of the
Access
Size (bytes) Notes
R
2
R
2
R
2
R
2
R
2
R
2
R
4
R
2
R
4
R
4
R
4
R
4
R
2
R
2
R
2
R
2
R
2
R
32
Modbus register maps
0.2–10.0 microns
0.2–10.0 microns
0.2–10.0 microns
0.2–10.0 microns
CAL
TMP
RH
FLO
YYYY (2000–9999)
MD (1–12, 1–31)
H (0–23)
MS (0–59, 0–59)
H (0-23)
MS (0–59, 0–59)
Bit wise mapped
0–4,294,967,295
0–4,294,967,295
0–4,294,967,295
0–4,294,967,295
mV
0.1 ºC external probe only
0.1% RH external probe only
mV
Double byte characters (16)