Modbus register maps
A.7 Sample mode parameters
Address
Register description
600
Number of count bins
601–616
Reserved
617
Count bin 1 limit
619
Count bin 2 limit
621
Count bin 3 limit
623
Count bin 4 limit
625–653
Reserved
654
ADC multiplier
655
DAC multiplier 1
656
DAC multiplier 2
657
DAC offset 1
658
DAC offset 2
659
DAC offset 3
660
DAC offset 4
661
Flow offset
662
ADC offset
663–699
Expansion
A.8 Diagnostic data
Address
Register description
700–705
Reserved
706
+5 VDC
707
+3.3 VDC
708
+5 VA
709–714
Reserved
715
Laser calibration
716
Laser current
717–723
Reserved
724
Error condition
725–749
Expansion
1
Set bits indicate a failure.
58
The sample mode parameters register
of a sample. Any updates to these registers will restart any active sample sequences.
Table 23 Sample mode parameters
Table 24
shows the Diagnostics data register that is updated at a 30 second (default) rate
or at the conclusion of any Test mode diagnostics.
Table 24 Diagnostics data record
Access
1
(Table
23) defines basic counting characteristics
Access
Size (bytes) Data format
R
2
R/W
4
R/W
4
R/W
4
R/W
4
R/P
2
R/P
2
R/P
2
R/P
2
R/P
2
R/P
2
R/P
2
R/P
2
R/P
2
Size (bytes) Data format
R
2
mV
R
2
mV
R
2
mV
R
2
mV
R
2
mA
R
2
System specific (e.g. sensor error)
1–4
0–4,294,967,295
0–4,294,967,295
0–4,294,967,295
0–4,294,967,295
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only
Factory calibration only