Chapter 3 Replacing Assemblies

Removing the Shroud

Removing the Shroud
Use the following procedure to remove the shroud:
WHEN THE SHROUD IS REMOVED FROM THE COUNTER,
DANGEROUS LINE VOLTAGES THAT CAN CAUSE SERIOUS INJURY
ARE EXPOSED. ALWAYS DISCONNECT THE POWER CORD BEFORE
REMOVING THE SHROUD.
Remove the front and rear bumpers (see "Removing the Bumpers" on
page 3-4).
The bumpers can be used to support and protect the Counter after the
shroud is removed. Place both bumpers on their inner edges on your work
surface, and then place the Counter on top of them. The rear bumper can
also serve as a holder for screws and other hardware temporarily removed
from the Counter.
Remove the handle and the handle mounts (see "Removing the Handle
and Handle Mounts" on page 3-5).
Remove the rear bezel (see "Removing the Rear Bezel" on page 3-6).
Use a #1 Phillips screwdriver to remove the three 4-40 x 3/16-inch
countersunk shroud screws from the top of the Counter near the front
edge of the shroud (see Figure 3-7).
3-8
Assembly Level Service Guide