Chapter 1 Performance Tests
Test 4: Channel 2 Power Measurement
Test 4e: Channel 2 22 GHz to 26.5 GHz Power
Measurement (0 dBm)
This test is appropriate for the Agilent 53151A and 53152A Counters.
Use the same test setup as described for Test 4a on page 1-25.
With the power output of the Microwave Amplifier still at 0 dBm
(adjust the power output of the Synthesizer, if necessary), set the
Synthesizer to the first frequency value listed for Test 4e in the
Performance Test Record on page 1-33 (22 GHz).
Record the actual power reading in the Performance Test Record.
Repeat steps 2 and 3 for the remaining frequency values listed for Test 4e.
Mark Pass or Fail for Test 4e in the Performance Test Record.
Test 4f: Channel 2 30 GHz to 46 GHz Power
Measurement (0 dBm)
This test is appropriate for the Agilent 53152A Counter only.
Use the same test setup as described for Test 4a on page 1-25.
With the power output of the Microwave Amplifier still at 0 dBm
(adjust the power output of the Synthesizer, if necessary), set the
Synthesizer to the first frequency value listed for Test 4f in the
Performance Test Record on page 1-33 (30 GHz).
Record the actual power reading in the Performance Test Record.
Repeat steps 2 and 3 for each of the remaining frequency values listed
for Test 4f.
Mark Pass or Fail for Test 4f in the Performance Test Record.