Chapter 6 Specifications
Measurement Specifications and Characteristics
Input Characteristics
Acquisition Time (1 MHz FM rate)
Channel 1
Channel 2 (FM Auto/FM Off)
Resolution
Channel 1/Channel 2
Emissions ("kickback noise")
Channel 1
Channel 2 (measuring/no input)
Residual Stability*
Channel 1
Channel 2
*Counter and source tied to same timebase
Accuracy
Channel 1/Channel 2
(LSD=Resolution selected)
Measurement Time
Channel 1
Channel 2
FM Tolerance
Channel 1
Channel 2 (FM Auto)
(FM Off)
AM Tolerance
Channel 1
Channel 2
Amplitude Discrimination
Channel 1
Channel 2
below 300 MHz
above 300 MHz
Power Measurement
Channel 1
Channel 2
Range
Accuracy at input connector**
(0 dBm to −20 dBm)
0.05–12.4 GHz
12.4–20 GHz
20–26.5 GHz
26.5–46 GHz
Resolution
Display
**see graphs for typical data
6-4
Agilent 53150A
N/A
125 ms/100 ms
1 Hz to 1 MHz
N/A
−40 dBm/<−70 dBm
N/A
0.6 LSD rms
±1 LSD
± residual stability
± timebase error x frequency
1/Resolution + 20 ms
1/Resolution +
Acquisition time + 20 ms
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
N/A
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
N/A
Counter sensitivity to +7 dBm
±1.5 dB
±1.5 dB
N/A
N/A
0.01 dB
dBm or milliwatts/microwatts
Agilent 53151A
N/A
125 ms/100 ms
1 Hz to 1 MHz
N/A
−40 dBm/<−70 dBm
N/A
0.8 LSD rms
±1 LSD
± residual stability
± timebase error x frequency
1/Resolution + 20 ms
1/Resolution +
Acquisition time + 20 ms
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
N/A
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
N/A
Counter sensitivity to +7 dBm
±1.5 dB
±1.5 dB
±2.0 dB
N/A
0.01 dB
dBm or milliwatts/microwatts
Assembly Level Service Guide
Agilent 53152A
N/A
140 ms/115 ms
1 Hz to 1 MHz
N/A
−40 dBm/<−70 dBm
N/A
1.25 LSD rms
±1 LSD
± residual stability
± timebase error x frequency
1/Resolution + 20 ms
1/Resolution +
Acquisition time + 20 ms
N/A
20 MHz p-p max to 26.5 GHz,
12 MHz p-p max above
26.5 GHz @ 10 MHz rate
1 MHz p-p @ 10 MHz rate
N/A
Any index provided minimum
signal level is not less
than sensitivity
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
N/A
Counter sensitivity to +7 dBm
±1.0 dB
±1.5 dB
±1.5 dB
±2.0 dB
0.01 dB
dBm or milliwatts/microwatts