Chapter 1 Performance Tests
Test 4: Channel 2 Power Measurement
Repeat steps 8 through 10 for each of the remaining frequency values
listed for Test 4a.
Mark Pass or Fail for Test 4a in the Performance Test Record.
Test 4b: Channel 2 22 GHz to 26.5 GHz Power
Measurement (–20 dBm)
This test is appropriate for the Agilent 53151A and 53152A Counters.
Use the same test setup as described for Test 4a on page 1-25.
With the power output of the Microwave Amplifier still at –20 dBm
(adjust the power output of the Agilent 83650B Synthesizer, if necessary),
set the Synthesizer to the first frequency value listed for Test 4b in the
Performance Test Record on page 1-32 (22 GHz).
Record the actual power reading in the Performance Test Record.
Repeat steps 2 and 3 for each of the remaining frequency values listed for
Test 4b.
Mark Pass or Fail for Test 4b in the Performance Test Record.
Assembly Level Service Guide
1-27