Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity
Test 2: Channel 1 Frequency Sensitivity
This set of tests verifies the frequency-sensitivity specifications of
Channel 1 of the Agilent 53150A/151A/152A Counters.
Equipment Required
Agilent 3325A/B Synthesizer
Agilent 83650B Synthesizer
Agilent 437B Power Meter
Agilent 8487A/D Power Sensor
Agilent 11730A Power Sensor Cable
Agilent 11667C Power Splitter
Agilent 8490D (Opt. 020) 20 dB Attenuator
Agilent 1250-2015 BNC (m) to SMA (f) Adapter
Agilent 11904A 2.92 mm (m) to 2.4 mm (m) Adapter
Agilent 10503 Series Coaxial Cable (BNC m to m)
2 mm (m) to 2 mm (f) Cable (2)
Test 2a: 10 Hz to 20 MHz Frequency Sensitivity
This test is appropriate for the Agilent 53150A, 53151A, and 53152A
NOTE
Counters.
1
Connect an Agilent 10503 Series Coaxial Cable between the RF output
connector on the Agilent 3325A/B Synthesizer and Counter's CHANNEL 1
connector.
1-16