Chapter 3 Replacing Assemblies

Removing the Handle and Handle Mounts

Removing the Handle and Handle Mounts
Use the following procedure to remove the carrying handle and handle mounts:
Rotate the handle to the vertical position above the Counter, pull the ends
of the handle outward at the mounts to detach it, and set it aside.
Figure 3-2. Removing the Handle
Remove the front bumper (see "Removing the Bumpers" on page 3-4).
Use a T15 TORX screwdriver to remove the two screws that attach each of
the handle mounts to the Counter, and remove the mounts.
TORX Screws
Figure 3-3. Removing the Handle Mounts
When reattaching the handle mounts, limit the torque on the TORX
screws to 4 to 5 inch-pounds.
Assembly Level Service Guide
Counter
Handle Mount
Handle Mount
TORX Screws