Remedy
Make sure the pump has a multi-drop address. Make sure that the commands are sent
using the multi‑drop protocol.
Fault
The green Normal LED does not light or the pump does not rotate at full speed or the
pump fails while in operation
Cause
The inlet pressure is too high.
Remedy
Decrease the pumping load or check for a gross leak in the system.
Cause
The pump is too hot while in operation.
• Increase the cooling of the pump. Change from air-cooling to water-cooling (refer
Remedy
to
Table: Pump performance data on page 30
cooling requirements).
• Increase cooling water flow or decrease the water temperature or do both.
• Check that external heat sources (such as system bakeout heaters) are not
excessive.
Cause
The rotor does not rotate freely.
Remedy
The pump bearings can be damaged. Contact us or the supplier.
Fault

Ultimate pressure cannot be reached

Cause
Pressure is limited by water vapour.
Remedy
Bake the system and pump.
Cause
The vacuum gauges are contaminated.
Remedy
Clean or replace the vacuum gauges.
Cause
Pumping speed is not sufficient (because of poor conductance between the pump
and the gauge or the chamber is too large).
Remedy
Increase the conductance or decrease the volume.
Cause
Inlet pressure is more than 0.5 mbar (50 Pa).
Remedy
If the interstage inlet pressure is too high, the inlet pressure at the turbomolecular inlet
is increased. Make sure that the interstage inlet pressure is less than 0.5 mbar (50 Pa).
Cause
The backing pressure is more than 10 mbar (1x10
Remedy
The backing pressure can be too high. Check for backing pipeline leaks. If the
throughput is high, a larger backing pump can be necessary.
Cause
The high vacuum area of the system is contaminated.
Remedy
Clean the high vacuum system.
Cause
Check the system for leaks and contamination.
Remedy
If a leak is found, repair it and clean the contamination.
Cause
The inlet pressure is less.
B8G000880_D - Fault finding
Page 87
for maximum inlet pressure and
3
Pa).