Calibration Procedure - Datacolor CHECK II Quick Start Manual

Table of Contents

Advertisement

Calibration Procedure

To start the calibration routine, do the following:
1.
2.
3.
4.
24  Instrument Calibration
Tap the Home icon on the graffiti pad, to display the
Datacolor CHECK II main menu.
TIP
The Calibrate option is only available from the Main
Menu.
Tap the Calibrate option. The following screen displays.
Cover the black trap with the port opening, and press the
Measure button. When it is finished, you are prompted to
place the white tile over the port.
NOTE
The serial number on the screen should match the serial
number of the white tile you are using.
Cover the white tile with the port opening, and press the
Measure button. When finished, you will see a message
that calibration was successful.
CHECK II Quick Start Guide