Routine Functions

Overview

This section covers instrument calibration, sample measurement, and
color evaluations.

Instrument Calibration

When calibrating, you must simulate the conditions under which you are
making measurements. If you are measuring with the stapler foot in
place, it must be attached during calibration, and the black trap and
white tile must be positioned against the bottom of the foot.
NOTES
When the calibration interval has fully elapsed, the instrument must be
recalibrated to make a measurement.
See Datacolor CHECK II User's Guide, for instructions to calibrate the
adjustable UV filter included with the Datacolor CHECK II PRO model.

Calibration Requirements

You will need the black trap and white tile to perform the
calibration routine.
If you are using the stapler foot assembly, you must attach the
stapler foot to the instrument before calibrating.
CHECK II Quick Start Guide
Overview  23