Troubleshooting guides

6.5.1
Problems with communications
Identify the cause
Recommended Actions
Device does not stay on segment
Steps to identify cause
Check segment
Recommended actions
6.5.2
Troubleshooting guide
Note
Use the sections below if other devices appear on the segment, communicate, and remain
on the segment. If other devices don't appear on the segment, communicate, or stay on
the segment, check the electrical characteristics of the segment.
If you have established communications, but you see a Block_err or an Alarm condition,
see
Problems with
recommended action (see
Device does not show up on segment
Potential cause
Unknown
Recommended action
Recycle power to the device.
Potential cause
No power to device
(2)
Consult your system integrator before performing the corrective action.
Reference Manual
1. Check wiring and power to device.
2. Recycle power to device.
3. If you can identify the problem, perform the Recommended Action in
does not show up on
4. If the problem persists, contact your local Emerson representative.
1. If you can identify the problem, perform the Recommended Action in
does not show up on
2. If the problem persists, contact your local Rosemount
communications. If you can identify the problem perform the
Recommended
segment.
segment.
(2)
actions).
Troubleshooting
July 2020
Device
Device
representative.