Trend arrow
CAUTION
If the app is not displaying a trend arrow, do not use your
GlucoMen Day CGM to make treatment decisions. Perform
a blood glucose test before making any treatment deci-
sions.
24
What it means
Your glucose level is rising at a rate of 0.19
mmol/L per minute or greater. You cannot cali-
brate if this arrow is visible.
Your glucose level is rising at 0.14-0.19 mmol/L
per minute. You cannot calibrate if this arrow is
visible.
Your glucose level is rising at 0.08-0.14
mmol/L per minute.
Your glucose level is changing less than 0.08
mmol/L per minute.
Your glucose level is falling at 0.08-0.14
mmol/L per minute.
Your glucose level is falling at 0.14-0.19 mmol/L
per minute. You cannot calibrate if this arrow is
visible.
Your glucose level is falling at a rate of 0.19
mmol/L per minute or greater. You cannot cali-
brate if this arrow is visible.
IMPORTANT: When using the GlucoMen Day CGM,
always carry a blood glucose meter.
6.3.3
Status icons
The status icons indicate the current status for calibration
and transmitter signal.
Calibration icons
The calibration
icons indicate
if it's time to
calibrate or if the
system cannot be
calibrated right
now.
Cannot calibrate right now.
Time to calibrate. Add a blood
glucose reading.
System is calibrated.

Advertisement