Setting The Maximum Skin Dose Rate; Prerequisites; Preparations - Siemens SIREMOBIL Compact L Service Instructions Manual

Table of Contents

Advertisement

Adjustments / Programs

Setting the maximum skin dose rate

WARNING
NOTE

Prerequisites

You must program one of the FL curves S2, HC2 or IODINE.
The FL curves S2, HC2 or IODINE should be used for the following adjustment.
The X-ray tube cover must be installed.
Use the dose ionization chamber of the dose measurement device to measure dose
rates greater than 10 R/min.
If you are using an area dose product measurement device, mount the dose measure-
ment chamber above the radiation exit port of the SIREPHOS.

Preparations

Attach the dose ionization chamber at a
distance of 30 cm from the I.I. housing
(Fig. 2).
Select one of the FL curves S2 / HC2 or
IODINE according to the programming.
(Push-key)
Briefly release fluoro.
Select kV stop mode and enter 110 kV
(5 mA).
NOTE
Siemens AG
Medical Solutions
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
Setting the maximum skin dose rate is necessary only for coun-
tries where DHHS regulations apply.
If the maximum tube current was already reduced, the mA value
displayed at 110 kV should be recorded and used for the calcula-
tion. The maximum skin dose rate to be set should equal 9 R/min.
SPR2-130.061.02
Rev. 11
5
30 cm
Chamber
Sirephos
Fig. 2
Page 9 of 30
09.04
CS PS 24
5 - 9
Input
screen
Holder
SIREMOBIL Compact
System Manual
5
5