5 - 6
Setting up needles
1.
Press the Needle Setup button in the service dialog.
Type
Fine
Core
Close
Fig. 7 Setting up needles dialog
2.
Press Add... .
You can only add needles which match the needle guides with the fixed
diameters: 0.7, 0.9, 1.2, 1.65, or 2.1 mm.
Enter needle values.
Type
L1
L2
OK
Fig. 8 Needle values dialog
3.
Select needle type (fine or core) from the option menu and enter needle length in
the text field.
L1 shall be 30 to 175 mm when selecting fine needle, L2 is not applicable for fine
needle. When selecting core, L1 shall be set to <170 mm.
Opdimar
Service
Needle Setup
L1 (mm)
L2 (mm)
90.5
-
110.0
5
Needles
Core
(mm)
(mm)
Cancel
Help
Register 5
SPB7-230.061.11
Rev. 02
Add...
Edit...
Delete
Help
Page 6 of 30
07.03
CS PS 24
Service mode
Siemens AG
Medical Solutions
5