Setting The Maximum Filament Frequency - Siemens POLYMOBIL III Service Instructions Manual

Hide thumbs Also See for POLYMOBIL III:
Table of Contents

Advertisement

Notes on trouble-shooting

Setting the maximum filament frequency

POLYMOBIL OFF
Remove fuse F5 on D920.
Connect oscilloscope on D910 to MP "CAL" and "GND".
POLYMOBIL ON
NOTICE
Using P3, set the maximum filament frequency on D910 (100 kHz
Siemens AG
Medical Solutions
Err 3 is displayed.
RXB8-115.061.02
Rev. 01
Page 15 of 24
10.95
CS PS 24
4 - 15
4
10 µs)
POLYMOBIL III