Setting The Preset Counter / Pulse Value (Imps) - TESTO 6448 Instruction Manual

Compressed air counter probe
Hide thumbs Also See for 6448:
Table of Contents

Advertisement

5 Operation
Representation
in display
rEF.P
rEF.T
5.1.5.

Setting the preset counter / pulse value (ImPS)

The instrument has seven adjustment ranges:
LED
Display
1
2
3
4
5
+
6
+
7
+
Setting procedure:
Press
Press and hold [Set].
The current numerical value flashes for 5 seconds, then the first
digit is active (digit flashes, can be changed).
Set the desired value as specified in the following table.
First select the desired adjustment range (1, 2, 3 etc.).
Enter the number from left (first digit) to right (fourth digit).
Press
38
Description
Standard pressure; measurement and display values for
flow correspond to this value.
Select
rEF.P
and set the desired standard pressure:
Adjustment range: 950...1050 hPa in 1 hPa increments
Standard: to 1013 hPa
Standard temperature; measurement and display values
for flow correspond to this value.
Select
rEF.T
and set the desired standard temperature:
Adjustment range: 0...25 °C in 1 °C increments
Standard 15 °C
Set
[Mode/Enter]
until
[Mode/ Enter]
briefly when all four digits are set.
Value
0.001 to 9999 m³
10.00 to 99.99 m³
100.0 to 999.9 m³
1000 to 9999 m³
10000 to 99990 m³
100000 to
999900 m³
1000000 m³
OU1
to ImP.
ImPS
is displayed.
Resolution
0.001 m³
0.01 m³
0.1 m³
1 m³
10 m³
100 m³