Representation
in display
FOUx (FOU1,
OU2)
dAP
rTo
Description
Behaviour of output x in the event of an internal error.
Adjustable parameters:
On = output x switches ON in the event of an error.
Off = output x switches OFF in the event of an error
(With both values (On, Off), the meter no longer runs in
the event of an error
OU = output 1 switches as defined with the parameters
irrespective of any error.
Measurement value damping / damping constant in
seconds
From 0.0 to 1.0 in 0.2 increments
dAP = 0; measuring value is completely undamped.
Resetting the quantity counter
After a set period of time, the counter is automatically
cleared and a new counting interval begins.
The following time spans can be set:
1h to 23h (reset after 1 to 23 hours)
1d to 6d (reset after 1 to 6 days)
1w to 8w (reset after 1 to 8 weeks)
Additionally:
OFF
= reset after counter overflow (32 bit number
overflow, i.e. overflow at 4,294,967.295 m³).
rES.T
= manual reset:
The counter is manually cleared and a new counting
interval begins:
Press
[Set]
until
rES.T
Now press
[Mode/Enter]
A manual reset can also be carried out when a time
period for an automatic reset has been set.
appears.
briefly.
5 Operation
35