222DS BENCHTOP SHAKING INCUBATOR
4.0 Safety Information
Before cleaning disconnect the device from mains power supply. Device could be cleaned with a soft,
lightly wet cloth. Don't use cleaners or liquids on aerosol basis.
Do not use the device nearby water sources. Take care, that water will not drop in the device, especially
by cleaning procedures.
Do not use device in aggressive atmosphere.
NOTE: In case that device is not functioning properly, even if you have exactly followed instructions
described in User's Manual, you are allowed to use only those commands and procedures, which are
allowed by User's Manual. Use of any other commands and procedure's adjustments could result in
device destruction or longer service repair time. If the equipment is used in a manner not specified by
the manufacturer, the protection provided by the equipment may be impaired.
Do not shake flammable or explosive samples.
When closing the lid, do not put your hands between the lid and housing, because they can
get crushed.
When the lid is open, do not put your fingers in the area between the housing and
rotating platform. Danger of mechanical injury.
The lid may be hot when the device is working at higher temperatures. Take care not to burn
your hands.
The back side of the housing may be hot when the device is working at higher temperatures.
Take care not to burn your hands.
Power supply cord shall be with nominal cross section area of at least 18 AWG.
5.0 Installation
5.1. Device Placement
When selecting the right place for device, please consider following:
Put the device on smooth, horizontal and stable place.
Leave enough space beyond the device for normal air circulation, min. 15 cm.
Leave enough space around the device, min. 15 cm, for easy access and maintenance.
Avoid locations subject to shocks and vibrations.
Don't use the device in surroundings, where there are fast temperature and humidity changes. Please
avoid also such places, which are under direct access of sunlight or places nearby heat producing devices.
Do not use the device in a flammable or explosive atmosphere.
NOTE: There has to be easy access to the shaker control panel and main plug in case of emergency.
5.2. Attaching the Power Cord
First connect the main power cord to the shaker then connect the plug end of the cord to a grounded
wall socket. Avoid lines on which powerful electric motors, refrigerators and similar devices are
connected.
Be careful when you plug the cord to a grounded wall socket.
Do not touch the plug with wet hands.
Do not pull the plug by the cord.
5.3. Attaching Shaking Platforms
The Shaking Incubator has a variety of platforms available, to meet most of the shaking needs. These
platforms mount to the shakers via four mounting platforms which easily plug into four rubber
mounting points on the top of the shaking incubator.
Need help?
Do you have a question about the 222DS and is the answer not in the manual?