1.0 Introduction The 222DS Benchtop Shaking Incubator is a combination product of a shaker and a temperature chamber, designed for shaking and heat treatment of sensitive samples. It replaces two devices, reduces time and space needed. It is very suitable for biochemistry, microbiology and medicine laboratories in which thermal and shaking treatment of samples is a daily used routine.
The Shaking Incubator has a variety of platforms available, to meet most of the shaking needs. These platforms mount to the shakers via four mounting platforms which easily plug into four rubber mounting points on the top of the shaking incubator. 222DS BENCHTOP SHAKING INCUBATOR...
6.0 Operating Instructions 6.1. Control Panel 88.8 TEMP START Encoder STOP POWER SPEED 88.8 TIME 1. POWER key – switch ON (illuminates when on) or OFF. 2. RUN – green signal light illuminated when shaking incubator is running. 3. TEMPERATURE – yellow signal light illuminated when the shaking incubator is set on temperature. 4.
Press encoder knob. Speed signal light blinks. With rotating encoder clockwise (+) or counter-clockwise (-), set rotating speed to desired value: 34 = 34 rpm 120 = 120 rpm 37.5 TEMP START Encoder STOP POWER SPEED TIME 222DS BENCHTOP SHAKING INCUBATOR...
Page 7
If you want to use the shaking incubator without shaking - set OFF on display. You get OFF function on display, when you rotate encoder under 20 RPM or above 300 RPM. Press encoder knob. Time signal light blinks. With rotating encoder clockwise (+) or counter-clockwise (-), set time to desired value from 1 min to 99 h: 70.3 = 70h 30 min 1.55 = 1h 55 min...
6.4 Shaking Without Heating If you want to shake the samples without heating them, set Temperature to OFF. 6.5 Heating Without Shaking If you want to heat the samples without shaking them, set Speed to OFF. 222DS BENCHTOP SHAKING INCUBATOR...
Set temperature is not reached in 2 hours. Shaking incubator is automatically stopped. Call for service. Should you have a question about the Labnet 222DS Shaking Incubator or required service for the unit, contact Corning Customer service at 800.492.1110 or 978.442.2200. Do not send a unit for service without first calling to obtain a repair authorization (RA) number.
Before you begin cleaning the device, unplug mains cord from power supply socket. Before using any other cleaning or decontamination methods, except those recommended by the manufacturer, you should check with the manufacturer, that the proposed method would not damage the equipment. 222DS BENCHTOP SHAKING INCUBATOR...
10.0 Positioning of Attachments on Shaking Platform You can mount different attachments on shaking platform. Follow below diagram as a guideline for positioning different flasks and accessories. 11.0 Accessories Product Cat. No. Qty/Pk Universal platform 11.8 x 11.8 in. (30 x 30 cm) I-5230-DS Petri dish shelf I-5220...
Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and electrical equipment (WEEE) provided with the products and available at www.corning.com/weee. 222DS BENCHTOP SHAKING INCUBATOR...
Page 13
To request certificates, please contact us at www.labnetlink.com. Warranty/Disclaimer: Unless otherwise specified, all products are for research use or general laboratory use only.* Not intended for use in diagnostic or therapeutic procedures. Not for use in humans. These products are not intended to mitigate the presence of microorganisms on surfaces or in the environment, where such organisms can be deleterious to humans or the environment.
Need help?
Do you have a question about the 222DS and is the answer not in the manual?
Questions and answers