If you want to use the shaking incubator without shaking - set OFF on display. You get OFF function on
display, when you rotate encoder under 20 RPM or above 300 RPM.
Press encoder knob.
Time signal light blinks. With rotating encoder clockwise (+) or counter-clockwise (-), set time to
desired value from 1 min to 99 h:
POWER
If you want to set Time to hold (continued operation) rotate encoder under 0.01 min or above 99.0 h to
get "HLd" on display.
Press START/STOP key once to exit the changing of parameters. You are now in starting mode.
Only signal light of the last changed parameter shines.
NOTE: You always change the parameter, whose signal light blinks - temperature, speed or time.
You can always change one, two or three parameters. To exit the changing of parameters, always
press START/STOP key once.
Press START/STOP key to start the operation.
Run signal light shines. Shaking incubator counts down the time from set value.
You can move between the values of time and speed during the operation, by pressing the encoder.
These two values are shown alternately on bottom display.
POWER
70.3 = 70h 30 min
1.55 = 1h 55 min
0.01 = 1 min
RUN
37.5
TEMP
SPEED
0.01
TIME
RUN
37.5
TEMP
SPEED
0.01
TIME
START
Encoder
STOP
START
Encoder
STOP