Chapter 2 Overview

Overview

Materials
Assembly
8
Thermal Cycler Temperature Verification System For GeneAmp
When you perform a temperature verification test on a 2720 thermal cycler, use the
materials in the kit (PN 4317939).
Figure 3
shows a properly assembled system. Note that the frame is positioned
around the wells with the channel facing you. The location tab on the frame should
align with the locator tab on the thermal cycler. The probe wire is threaded through
the channel to prevent damaging the wire when you close the thermal cycler heated
cover.
Figure 3 Proper assembly of the Temperature Verification System for 2720
thermal cycler
Figure 4 Temperature verification frame
®
PCR Systems 2720, 9600, 9700, 9800,
®
and Veriti
System User Guide