Sign In
Upload
Download
Table of Contents
Contents
Add to my manuals
Delete from my manuals
Share
URL of this page:
HTML Link:
Bookmark this page
Add
Manual will be automatically added to "My Manuals"
Print this page
×
Bookmark added
×
Added to my manuals
Manuals
Brands
Applied Biosystems Manuals
Laboratory Equipment
2720
User manual
Thermal Cycler Temperature Verification System For Geneamp ® Pcr Systems 2720, 9600, 9700, 9800; Data Sheet: Temperature Non-Uniformity Test - Applied Biosystems GeneAmp 2720 User Manual
Thermal cycler temperature verification system
Hide thumbs
Also See for GeneAmp 2720
:
User manual
(122 pages)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
page
of
151
Go
/
151
Contents
Table of Contents
Bookmarks
Table of Contents
Advertisement
Chapter 4 Data Sheet: Temperature Non-Uniformity Test
Data Sheet: Temperature Non-Uniformity Test
48
Thermal Cycler Temperature Verification System For GeneAmp
When you run the temperature non-uniformity test, record the setpoint values for the
wells L1 to L6 and R1 to R6 listed on this data sheet. At the end of the test, compare
the values that are displayed on the thermal cycler with the values that are recorded
here.
You can photocopy this page.
Note:
Date
Tested by
Probe serial no.
Thermometer
serial No.
Setpoint value
L1
L2
L3
L4
L5
L6
R1
R2
R3
R4
R5
R6
95 °C
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
55 °C
®
System User Guide
Table of
Contents
Previous
Page
Next
Page
1
...
53
54
55
56
57
58
59
60
Advertisement
Chapters
Table of Contents
3
Introduction 1
9
Temperature Verification Tests for the 2
15
Temperature Verification Tests for the 3
29
Temperature Verification Tests for the 4
43
® Temperature Verification Tests for the Geneamp 9700, Dual 384-Well Sample Block Module 5
57
® Temperature Verification Tests for the Geneamp 9700, Auto-Lid Sample Block Module 6
71
® Temperature Verification Tests for the Geneamp 9700, 0.5-Ml Sample Block Module 7
85
Temperature Verification Tests for the Geneamp 9800 8
99
Temperature Verification Tests for the Geneamp 9600 9
113
Temperature Verification Tests for the Veriti Thermal Cycler 10
125
Table of Contents
Related Manuals for Applied Biosystems GeneAmp 2720
Laboratory Equipment Applied Biosystems 2720 User Manual
Thermal cycler for amplification of nucleic acids (122 pages)
Laboratory Equipment Applied Biosystems GeneAmp PCR System 9600 User Manual
(296 pages)
Control Unit Applied Biosystems GeneAmp PCR System 9700 User Manual
Base module (132 pages)
Laboratory Equipment Applied Biosystems GeneAmp PCR System 9700 User Manual
Base module (128 pages)
Laboratory Equipment Applied Biosystems GeneAmp PCR System 9700 User Manual
Dual 96 sample block module (53 pages)
Laboratory Equipment Applied Biosystems 7900HT User Manual
Fast real-time pcr system and sds enterprise database (384 pages)
Laboratory Equipment Applied Biosystems 7500 Getting Started Manual
Fast real-time pcr system (106 pages)
Laboratory Equipment Applied Biosystems ViiA 7 Real Quick Reference Manual
Time pcr system (23 pages)
Laboratory Equipment Applied Biosystems 7300 Installation And Maintenance Manual
Real-time pcr system (174 pages)
Laboratory Equipment Applied Biosystems QuantStudio 3D Quick Reference
Real-time pcr system (10 pages)
Laboratory Equipment Applied Biosystems QuantStudio 3 Site Preparation Manual
Real-time pcr systems (18 pages)
Laboratory Equipment Applied Biosystems 7300 Getting Started Manual
Fast real-time pcr system (78 pages)
Laboratory Equipment Applied Biosystems QuantStudio 6 Flex User Manual
(194 pages)
Laboratory Equipment Applied Biosystems 7500 Getting Started
Real-time pcr system (152 pages)
Laboratory Equipment Applied Biosystems CytoFluor 4000 Series User Manual
Fluorescence multi-well plate reader (221 pages)
Laboratory Equipment Applied Biosystems A30207 User Manual
Automated thermal cycler (atc) system (70 pages)
Related Content for Applied Biosystems GeneAmp 2720
GeneAmp PCR System 9600 Sample Conversion Of Hold Times For Up Ramps (Dna Thermal Cycler To Geneamp Pcr System 9600)
Applied Biosystems GeneAmp PCR System 9600
GeneAmp PCR System 9700 How The 9700 Differs From The Geneamp Pcr System 9600
Applied Biosystems GeneAmp PCR System 9700
2720 Introducing The Applied Biosystems 2720 Thermal Cycler
Applied Biosystems 2720
A30207 Set Up The Automated Thermal Cycler (Atc) System On A Robotic System
Applied Biosystems A30207
QuantStudio 6 Flex Differences Between The Quantstudio ™ 6 And 7 Flex Real-Time Pcr Systems
Applied Biosystems QuantStudio 6 Flex
StepOne Real-Time Pcr System Software
Applied Biosystems StepOne
7300 Applied Biosystems 7300/7500/7500 Fast Real-Time Pcr System Installation And Maintenance Guide
Applied Biosystems 7300
7900HT Applied Biosystems 7900Ht Fast Real-Time Pcr System And Sds Enterprise Database User Guide
Applied Biosystems 7900HT
CytoFluor 4000 Series Controlling The Temperature
Applied Biosystems CytoFluor 4000 Series
7500 About The 7500/7500 Fast System
Applied Biosystems 7500
7300 About The 7300/7500/7500 Fast System
Applied Biosystems 7300
7500 Selecting The Pcr Method
Applied Biosystems 7500
GeneAmp PCR System 9700 Performing The Calibration Verification Test
Applied Biosystems GeneAmp PCR System 9700
ViiA 7 Real Calibrate The Viia 7 System
Applied Biosystems ViiA 7 Real
This manual is also suitable for:
Geneamp 9600
Geneamp 9700
Geneamp 9800
Table of Contents
Print
Rename the bookmark
Delete bookmark?
Delete from my manuals?
Login
Sign In
OR
Sign in with Facebook
Sign in with Google
Upload manual
Upload from disk
Upload from URL