5 - 4
Connecting the stand cable-harness
High-voltage connectors, alternative types
Type A
Fig. 4
High-voltage connector, alternative types
NOTICE
Mammomat 1000/3000
Installation and Start-Up
3
1
2
Type B
The high-voltage connector (1) is factory-adjusted for correct fit.
Mount the angular cable-sleeve halves and sleeve nut (Type B
only). The high-voltage connector shall be inserted with silicone
oil, silicone-rubber ring (2) and silicone-rubber disc (3). Lubricate
the silicone-rubber disc and the pins with silicon oil before
mounting the disc. Make sure that there aren't any air bubbles
between the disc and the connector.
Register 3
SP B7-230.033.07
Rev. 03
Cable connections
3
1
2
Page 4 of 14
11.99
SPS-UD
5
5
Siemens-Elema AB
Solna, Sweden