Attaching the swivel-arm covers

Attaching the swivel-arm covers

Arranging the swivel-arm system

1.
2.
3.
4.
5.
6.
7.
8.

Connecting the cables to control-button boards and patient handles

MAMMOMAT 1000/3000
Fig. 1 Connecting the cables
Before installing the side covers, the cables are to be connected.
MAMMOMAT 1000/3000
1.
2.
Siemens-Elema AB
Solna, Sweden
Remove the cable tie securing the balancing spring at the back of the stand.
Install object table, no compression force (swivel-arm system still upside down).
Remove the protective strips from the metal curtain.
Run the lifting carriage upwards.
Put back the protective strips onto the metal curtain.
Rotate the swivel-arm system to 0 degrees by pressing one of the switches for
clockwise rotation on the supporting arm.
System and mains "OFF".
Remove the swivel-arm system transport safeguard, see Chapter 3 "Preparatory
work".
1
2
2
1
Connect cables X807 (1) to control-button circuit boards D807 (2).
Connect ground wires (3) to patient handles (4).
Register 3
4
3
3
4
SP B7-230.033.07
Rev. 03
11.99
7
Page 1 of 2
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
7 - 1
7
7
7