Checking And Programming With The Service Pc; Sensitivity - Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

Start-up and functional test of the IONTOMAT

Checking and programming with the service PC

Use the service PC to check and program the IONTOMAT.
Operation of the service PC is described in Chapter 20 "Appendix".
The tables in this kit are optimized for the following four film/screen combinations.
Table 1
Kodak Min-RE
Fuji UM-MA HC
Agfa Mamoray MR6
Dupont Microvision C
Program the IONTOMAT values as follows:
NOTICE

Sensitivity

Start with the sensitivity listed in the table for the film/screen combination used by the cus-
tomer. Program this values for "wing 1" and "wing 2" under H. If the customer uses one
more film/screen combination, program this value for "wing 1" and "wing 2" under D, oth-
erwise program the same values as under H.
If the film/screen combination used by the customer is not listed in table 1, start
with sensitivity 11.
NOTICE
IONTOMAT PM - Sensitivity
1 Help
Fig. 1
Values from 0 to 31 can be entered.
Siemens-Elema AB
Solna, Sweden
Film
Screen
Kodak Min-R
Fuji HR Mammo Fine
Agfa MR Detail S
Dupont Quantavision ME Dupont Quantavision 12
If it is a M 3000 stand equipped with one "wing" only and counter-
weight, the programming for "wing 2" shall be omitted. If it is a
M 1000 stand, the programming is done for "base plate".
Be sure to use the same type of cassette, screen and film in each
object table as the customer uses.
Wing 1
Wing 2
Valid entries from 0 to 31 in 1/2 E.P.
Input of decimal point not required!
2 Save
3
Register 3
Cassette
Kodak Min-R2
Fuji EC-MA
Agfa MR Detail S
H
11.0
11.0
4
5
6
SP B7-230.033.07
Rev. 03
11.99
11
Sensi-
Correction
tivity
curve
11
9.5
7
D
11.0
11.0
7
8
9
Page 1 of 12
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
11 - 1
11
A
B
C
D
11
10 Quit

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat 3000

Table of Contents