Main Voltage Connection - Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

5 - 10
6.
Fig. 12 Connecting the decompression button cable at the stand

Main voltage connection

The equipment has been set to 400 V, 2-phase, in the factory (see test certificate).
Adjustment to the line frequency is not needed. If the mains voltage is not 400 V, 2-phase
at the installation site, a reconnection of the mains is necessary.
Should a reconnection for the mains voltage be necessary, this must be done at trans-
former T10 as well as p.c. board D711 (insertion or removal of fuses and labeling is spec-
ified in the wiring diagram, see Line Input page 5-3).
Change the rated voltage labels to the correct voltage (there are two labels on the gener-
ator and one on the stand). Labels are included in the installation material.
NOTICE
Mammomat 1000/3000
Installation and Start-Up
Shield the decompression button cable to the stand chassis as shown in Fig. 12.
Connect to terminals 2 and 3 in terminal block X803 on the inside of the stand
chassis, see Fig. 12.
For more details about changing mains, insertion or removal of
jumpers, fuses, labeling etc, see Wiring Diagram.
Register 3
1
2
SP B7-230.033.07
Rev. 03
11.99
Cable connections
X885
1
2
3
4
X880
Page 10 of 14
Siemens-Elema AB
SPS-UD
5
Solna, Sweden