The adjustment lever marks colours to indicate the functional state
of the orthotic joint
Col
Illustration
our
Red
see fig. 3
Yel
see fig. 4
low
Gre
see fig. 5
en
Finishing the orthosis
Required materials: 636W28* special adhesive, degreasing cleaner
The bars must be glued in the insertion zones to ensure stability.
1) Clean the connecting surfaces with a degreasing cleaner.
2) Apply the special adhesive to the connecting surfaces.
3) Insert the bars.
4) Secure with the enclosed screws.
5) Allow to cure for at least 4 hours.
INFORMATION: Final bonding strength will be reached after
16 hours.
6 Cleaning
Promptly clean the joint after contact with water containing salt, chlorine or
soap, or if it gets dirty.
1) Rinse the orthotic joints with clean fresh water.
2) Dry the orthotic joints with a cloth or allow them to air dry. Avoid expos
ure to direct heat (e.g. from an oven or radiator).
7 Maintenance
INFORMATION
The product may be exposed to increased loads by the patient.
► Shorten the maintenance intervals according to the expected loads.
The manufacturer recommends a semi-annual inspection to verify functional
ity and check for wear and tear to the orthotic joints.
Spare parts are listed in the section "Components/design".
Only use 633F7 special lubricant.
Functional state
The orthotic joint is locked.
The orthotic joint can be moved in the flexion direc
tion and is limited by the extension stop.
The orthotic joint is unlocked in flexion and exten
sion.
17