Otto Bock LP2-00 LP Symes Instructions For Use Manual page 15

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 12
Quantity
Designation
1
Protective sock (black)
1
Heel wedge set
1
Lamination adapter set (including two
screws)
Additional accessories/spare parts (not included in the scope of delivery)
Designation
Footshell (without cap)
Tool to change the footshell
Protective sock (white)
5 Preparing the product for use
CAUTION
Incorrect alignment, assembly or adjustment
Injury due to incorrectly installed or adjusted as well as damaged prosthet­
ic components
► Observe the alignment, assembly and adjustment instructions.
NOTICE
Grinding the prosthetic foot or footshell
Premature wear resulting from damage to the product
► Do not grind the prosthetic foot or footshell.
5.1 Alignment
The prosthetic foot is connected to the prosthetic socket using a lamination
adapter. Subsequent alignment corrections are not possible. Therefore, the
prosthesis should be aligned as follows:
1) Fabricate a check socket.
2) Connect the prosthetic foot to the check socket.
→ Use filler between the check socket and the lamination adapter to
adjust the alignment and the installation dimension.
→ Secure the alignment by wrapping the connection with reinforce­
ment material.
3) Pull on the footshell (see page 16).
4) Check the static alignment and perform a dynamic fitting (see page 16,
see page 16).
→ The load line should run along the front edge of the lamination
adapter.
Reference number
S0-NPS-200*
KIT-00-11*
KIT-00-16200-00
Reference number
FTC-3M-1*
ACC-00-10300-00
SL=Spectra-Sock