Driving data
Climbing ability (basic model)**
Dynamic stability – uphill***
Maximum obstacle height
Driving distance range (on level sur­
faces)***
Braking distance (according to DIN EN
12184)****
* The specified speed can vary by ±10%.
** The control device and the motors must be protected against overloading. Therefore, the continuous climbing
ability depends on the overall weight (wheelchair weight + user weight + load), as well as the ground conditions,
exterior temperature, battery voltage and driving style of the user. In individual cases, the continuous climbing abil­
ity may be significantly lower than the value specified.
*** The specified driving distance range was determined under defined conditions according to ISO 7176-4. In
practice, the driving distance range can be reduced by up to 50 per cent. For information on this, see the section
"Driving distance range" in the instructions for use (user).
**** The braking distance can be correspondingly longer due to user weight, installed options and condition of the
tyres as well as weather and surface conditions.
Allowable environmental conditions
Operating temperature
Transport and storage temperature
Relative humidity
Corrosion protection
Corrosion protection
12 Appendices
12.1 Required tools
The following tools are required for adjustments and maintenance work:
Allen wrenches in sizes 3, 4, 5, 6
Ring and open-ended wrenches in sizes 10, 11, 13, 19 and 24
Torque wrench (measurement range 5-50 Nm)
12.2 Torque values of the screw connections
Unless otherwise specified, screw connections are tightened with the following torque values:
Thread diameter M4: 3 Nm
Thread diameter M5: 5 Nm
Thread diameter M6: 10 Nm
Thread diameter M8: 25 Nm
20
Climbing ability* without stand-up function (sitting): max. 10° (17%)
With activated stand-up function: max. 3° (5%)
10° (17%)
75 mm
Approx. 35 km
At 6 km/h [3.7 mph]: 1000 mm (on level surfaces); 2000 mm (incline)
At 10.0 km/h [6.2 mph]: 2100 mm (on level surfaces); 4200 mm
(incline)
-15 °C to +40 °C (+5 °F to +104 °F)
-15 °C to +40 °C (+5 °F to +104 °F)
45% to 85%; non-condensing
Coated frame