Spo2 Specifications - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008
in automatic mode
Typical
measurement
time
Normal mode
measuring range
(mmHg)
Measurement accuracy
Resolution
Initial inflation pressure
Overpressure
protection point
(software)
Overpressure
protection point
(hardware)
Pressure accuracy
Electrical characteristics
Supply voltage
Maximum power
consumption
Quiescent current
Maximum current
during measurement
Maximum current
during inflation
A.3.3 SpO2 Specifications
Standards compliant
Display range
SpO2 display
resolution
SpO2 checking
accuracy
20~40s
Adult
Systolic blood
40-270
pressure
Mean blood
20-230
pressure
Diastolic blood
10-210
pressure
Maximum average error: ± 5mmHg
Maximum standard deviation: 8mmHg
1mmHg
Adult
Pediatric
Adult: 300mmHg
Pediatric: 240mmHg
Adult: 320~330mmHg
Pediatric: 265~275mmHg
± 3mmHg
10V~14V DC
3.6w
50mA
180mA
300mA
ISO 80601-2-61
0%~100%
1%
2% (70%~100%); not define when lower than 70%
Pediatric
40-230
20-165
10-145
Default
150mmHg
100mmHg
版本: 1.0
Pressure setting range
80~240mmHg
80~200mmHg
- 84 -

Advertisement

Table of Contents
loading

Related Products for Northern Meditec Taurus E