Set [Min Alarm Vol] to [0];
[Vol.Set] smart hotkey → [Volume Setup] menu;
Select
Set [Alarm Volume] to [0].
When the alarm sound is turned off, the alarm state area on the screen shows the
If [Min Alarm Vol] is larger than 0, the system will cancel alarm sound off state.

7.4.3 Alarm Pause

Press the
[Pause] smart hotkey to temporarily stop the alarm of the monitor in the
following steps:
Alarm state icon area will display the
The light alarm and audible alarm of the physiological alarms are suspended, and the
alarm information is not displayed.
The remaining time of alarm pause is displayed in the physiological alarm area.
Alarm parameters and upper / lower limit stop flashing.
The audible alarm and light alarm of technical alarms are suspended, but the alarm
message is still displayed.
After the alarm pause is finished, the monitor will automatically cancel the alarm pause state.
During the alarm pause, you can also press
the alarm pause.
You can set the alarm pause time as follows:
Select
Select [User Maintenance >>] →enter the password and confirm → [User Maintenance]
menu;
Select [Alarm Param >>] → [Ala.Para Setup] menu;
Select [Alarm Pause Time], and set the alarm pause time as needed:
[1min] / [2min] / [3min] / [4min] / [5min] / [10min] / [15min]: Set the alarm pause
time to 1 min, 2 min, 3 min, 4 min, 5 min, 10 min or 15 min. By default, the alarm
pause time is 2 minutes.
[Permanent]: Set the alarm pause time to permanent.

7.4.4 Alarm off

[Menu] Smart Hotkey→[Main Menu];
icon.
[Pause] smart hotkey to manually cancel