Chapter5 Patient Information Management

Connect the patient to the monitor, and the monitor will display and store the physiological
data of the patient, so the patient can be monitored without admitting the patient. However,
admitting the patient correctly is very important.
If the monitor has admitted the patient, it is recommended to operate the monitor to discharge
the current patient before connecting to (not admitted) the next patient. Otherwise, the data of
the previous patient will be stored in the data of the current patient.
Warning
Whether the patient is admitted or not, the system will give a default value to
[Patient Cat.] and [Paced], and the user must confirm that the default value is
appropriate for the patient being monitored.
For patients with pacemakers, [Paced] must be set to [Yes]. Otherwise, the
pacing pulse will be treated as normal QRS wave group, and the system is
unable to detect the alarm status of [ECG Signal weak].
For patients without a pacemaker, [Paced] must be set to [No]. Otherwise, the
system is unable to detect the arrhythmias (including PVCs count) related to
ventricular premature beats, and fails to perform ST segment analysis.
5.1

Patient Setup Menu

You can manage the patients through the [Patient Setup] menu ; to enter the [Patient Setup]
menu, operate as follows:
[Menu] Smart Hotkey→ [Main Menu];
Select
Select [Patient Setup >>] → [Patient Setup] menu, as shown in Fig. 4-1.