Place the electrode on the patient.
Connect the ECG cable and ECG interface.
Warning
Check if the lead is normal before monitoring. When the ECG cable is
unplugged, the screen will display [ECG Lead Off] prompt, and trigger an
audible and visual alarm.
12.3.2 Selecting Lead
Select the ECG parameter area or wave area → [ECGSetup] menu;
Select [Other Setup >>] → [ECG Other Setup] menu;
Select [Lead Type], and select the ECG lead as needed:
[3-Lead]: 3-lead; ECG wave options: I, II, III.
[5-Lead]: 5-lead; ECG wave options: I, II, III, AVR, AVL, AVF, V.
12.3.3 Lead Name and Corresponding Color
The lead names in European standard and U.S. standard (represented with R, L, N, F, C in
European Standard, and represented with RA, LA, RL, LL, V in the U.S. standard) are shown
in Table 7 -1.
Table 7-1: Lead Name in European Standard and American Standard
European Standard (IEC)
Lead Name
R
L
F
N
C
Color
Lead Name
Red
Yellow
Green
Black
White
American Standard (AHA)
Color
RA
White
LA
Black
LL
Red
RL
Green
V
Brown