Disinfection; Alarm Setup - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008
2.
Sprinkle cleaning agent on NIBP cuff, connector and extension tube, keep 1 minutes for
the dry stains.
3.
Use a soft cloth to wipe smooth face. Use soft hair brush to brush visible stain and
irregular surface
Note
Please be especially careful to clean the air ball and control valve of whole air
system. Do not allow any liquid entering into reversing valve and saturated valve
Don't use a soft cotton ball and fiber to clean this accessory, because they will stick
on the cuff and extension tub
10.10.2

Disinfection

1.
Sprinkle bleach solution (Formula: the proportion of water and bleaching powder to
1:10) then keep 5 minutes
2.
Wipe off excess bleach solution and elute with distilled water again
3.
Natural dry cuff

10.11 Alarm Setup

In [NIBP Setup] menu, select [Alarm Setup >>] to enter [Alarm Setup] interface, and set
NIBP alarm switch, alarm level, upper and lower alarm limit. See 6.5 Alarm Setup for
detailed setting method.
版本: 1.0
- 54 -