Chapter14 Caring And Cleaning; Overview; Cleaning - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008

Chapter14 Caring and Cleaning

14.1 Overview

In the using process, please make sure that there is no dust on or near your device. To prevent
damage, please use the diluted detergents and disinfectants specified in this Manual, and use
the lowest possible concentration. For the damage or accident caused by using other materials
or methods, Northern does not assume any responsibility.

14.2 Cleaning

The device should be cleaned regularly. In the heavily polluted environment, increase the
frequency of cleaning. Before cleaning, please consult the hospital about device cleaning
requirements.
Below are available cleaning agents:
Diluted ammonia
Diluted sodium hypochlorite (washing bleach)
Diluted formaldehyde
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
Before cleaning:
1. Turn off the monitor, disconnect the power cord and remove the battery.
2. Use a soft cotton ball to adsorb appropriate amount of cleaning agent and wipe the
display screen.
3. Use a soft lint-free cloth to adsorb appropriate amount of cleaning agent and wipe the
surface of the device.
4. If necessary, use a clean, dry, lint-free cloth to remove any excess detergent.
5. Dry the device naturally in a ventilated cool environment.
Warning
版本: 1.0
- 71 -

Advertisement