HR
Measuring range
Resolution
Heart rate measurement
error
TEMP
Standards compliant
Measurement method
Measuring range
Resolution
Measurement accuracy
Number of channels
Alarm specifications
T1/T2 upper limit
T1/T2 lower limit
TD upper limit
RESP
Measurement method
Measuring lead
Wave gain
Respiratory impedance
range
Baseline impedance
Gain
Scan speed
A.3.2 NIBP Specifications
Standards compliant
Measurement method
Operating mode
Useful life
Measurement
interval 1/2/3/4/5/10/15/30/60/90/120/180/240/480min
3/5-lead monitoring
1bpm
± 1bpm or ± 1%, whichever is greater
EN12470-4, ISO 80601-2-56
Thermistor
5~50° C (41~122° F)
0.1° C
± 0.1° C
Two
Range
0.1° C~50.0° C
0° C~49.9° C
0~50° C
Thoracic electrical bioimpedance method
Lead I, II
× 0.25, × 0.5, × 1, × 2
0.5-5Ω
500-4000Ω
10 grades
6.25mm/s, 12.5 mm/s, 25mm/s
EN 60601-2-30/IEC 60601-2-30, EN 1060-1, EN 1060-3, EN1060-4
EN/IEC 60601-1.
Automatic oscillometric method
Manual, automatic, continuous
100, 000 times
Adult: 15bpm~300bpm
Pediatric: 15bpm~350bpm
Step
0.1° C
0.1° F