Latch Alarm - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008
Select [Default], and restore the alarm setup to the factory default.
Note
When setting the upper and lower alarm limit, confirm the patient category to be
adults or pediatric, and set its scope according to the clinical need. If the setting
exceeds the alarm limit, the alarm system will fail easily.
When the alarm limit is turned on, and the upper and lower alarm limits are
manually set, the instrument will display the upper and lower alarm limits
continuously, and the initial alarm preset value won't be provided additionally.
7.6

Latch Alarm

Physiological alarms can be set to [Latching] or [No latching].
[Latching]: Even if the cause of physiological alarm is cleared, the system will still be
"latched", that is, continue to display the alarm information corresponding to
physiological alarm, the alarm sound also continues, but the alarm mode has the
following changes:
 Parameters and upper or lower alarm limit are no longer flashing.
 Display the time that the latest alarm was triggered after the alarm message in the
physiological alarm area.
[No latching]: After the causes of physiological alarm are cleared, the system will no
longer prompt the physiological alarm.
The default alarm of the system is non-latching alarm; you can set the alarm as latching or
non-latching in the following steps:
Select
Select [User Maintenance >>] →enter the password and confirm → [User Maintenance]
menu;
Select [Alarm Param >>] → [Ala.Param Setup] menu;
Select [Latching Alarms], and set the alarm as needed:
[Latching]: Latching alarm.
[Menu] Smart Hotkey→[Main Menu];
版本: 1.0
- 38 -