Select [Alarm Delay], and set the alarm delay time as needed:
[Off]: Turn off the alarm delay.
[1s] / [2s] / [3s] / [4s] / [5s] / [6s] / [7s] / [8s]: Alarm delay time is 1 sec, 2 sec, 3
sec, 4 sec, 5 sec, 6 sec, 7 sec or 8 sec. By default, the alarm delay time is 4 seconds.
Select [ST Alarm Delay], and set the ST alarm delay time as needed
[Off]: ST alarm delay is off
[10s] / [20s] / [30s] / [45s] / [1min] / [2min] / [3min]: ST alarm delay time is 10
sec, 20 sec, 30 sec, 45 sec, 1 min, 2 min or 3 min. By default, the ST alarm delay
time is 20 seconds.
The system sets physiological alarm delay time and also sets 1~2s delay time after technical
alarm been triggered. Normally the delay time is limited to 5s.

7.5.3 Setting a Parameter Alarm

With SpO2 for example, select the SpO2 parameter area, select [Alarm Setup >>] in the
popup [SpO2 Setup] menu to enter the SpO2 alarm setup interface.
1. Turn on / off alarm
Select [Alarm Switch] and set the alarm switch as follows:
[On]: Turn on SpO2 alarm; when the parameter alarm occurs, the monitor will
prompt according to the set alarm level.
[OFF]: Turn off SpO2 alarm;
monitor won't prompt the parameter alarm.
2. Set the alarm level
Select [Alarm Level], and set the alarm level as follows:
[Low]: Set the alarm level to low.
[Mid]: Set the alarm level to medium.
[High]: Set the alarm level to high.
3. Set the alarm limit
In any cases, the alarm system only allows setting the values within the effective range of the
system, and the upper alarm limit must be higher than the lower alarm limit.
Select [Spo2 Low Limit] and set the lower limit of SpO2 alarm.
Select [Spo2 High Limit] and set the upper limit of SpO2 alarm.
4. Restore default alarm setup
icon is displayed in the parameter area, and the
版本: 1.0
- 37 -