A.3 Measurement Specifications
A.3.1 ECG/TEMP/RESP Specifications
ECG Specifications
Standards compliant
Lead type
Display sensitivity
Wave sweep speed
Bandwidth
CMRR
Notch
Differential
input
impedance
Electrode polarization
voltage range
Baseline recovery time
Calibration signal
Pacing pulse
Pulse identification
Average HR
Interval of HR
refreshing
HR change response
time
Tall T-wave
suppression
Alarm specifications
HR upper limit
HR lower limit
EN 60601-2-27/IEC 60601-2-27, GB 9706.25, IEC60601-2-25
3-lead
5-lead
2.5mm/mV (× 0.25), 5mm/mV (× 0.5), 10mm/mV (× 1.0), 20mm/mV
(× 2.0)
6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Diagnostic mode
Monitor mode
Surgery mode
Strong filter mode
>100dB
50/60Hz notch filter can be set to on or off
>5MΩ
± 400mV
<3s after defibrillation (in monitor and surgery mode)
1mV (peak - peak), accuracy ± 3%
For PACE pulses that meet the criteria below, PAEC will be marked on
the screen:
Detection range: ± 4mv ~ ± 700mv
Pulse width: 0.2ms ~ 2.0ms
Calculate from 15s data
Calculate once every second
Time from 80bpm to 120bpm: ≤ 10sec
Time from 80bpm to 40bpm: ≤ 10sec
For T-wave with 100ms QRS wave, 350ms QT period, 180ms duration
and 1.2mV amplitude, the HR calculation won't be affected
Range (bpm)
Adult: 16~300
Pediatric: 16~350
Adult: 15~299
Pediatric: 15~349
I, II, III
I, II, III, aVR, aVL, aVF, V
0.05Hz~100Hz
0.5Hz~40Hz
1Hz~20Hz
5Hz~20Hz
版本: 1.0
Step (bpm)
1